Valsteam ADCA
Спецификации
-
Applications
Temperature measurement in pipelines and vessels.
-
Max. operating conditions
36,5 bar @ +400 ºC
-
Temperature range
0 to +200 ºC (others programmable),
The device is calibrated from factory for a 0 to 200 ⁰C -
Temperature limits*
-10 to +300 ⁰C,
(*)The range must always fall between the device's temperature limits! -
Ambient temperature
-20 ºC to +80 ºC
-
Sensing element
RTD three wire device according to EN 60751 Class A
-
Enclosure/head
KNE
-
IP rating
IP 66
-
Output
4 to 20 mA (2-wire)
-
Transmitter
2 wire 4 to 20mA temperature transmitter
-
Transmiter thermal drift deviation
Input 0,06 ºC / 10 ºC ,
Output 0,02% of span / 10 ºC -
Power supply
10 to 30 V DC
-
Max. loop resistance
681 Ω at 24 V DC, 954 Ω at 30 V DC
-
Approvals
ATEX approved version on request
-
Sizes and connections
Male thread ISO 228 G – 1/2''